The 8 cubic meter air purifier environmental chamber is mainly used to measure the purification performance of the air purifier. The indicators are the purification ability of particulate matter and the removal rate of harmful gases. The test chamber can detect the natural decay of particulate matter and gas test of the air purifier.
SPS-KACA002-132:2018 "Indoor Air Purifier"
Other requirements:
Test environment: temperature: 23±ºC, humidity: 55±15%
Test cabin: 0.1m³/min<P<1.6m³/min: 8.0±0 5m³ cabin
1.6m³/min≤P<16m³/min: 29.5±1m³ cabin
P>16.0m³/min: (50.0±1)m³ cabin
Functional test of air purifier performance test chamber:
The air purifier test chamber is used to evaluate the purification efficiency, purification level, purification life and other technical indicators of the air purifier. The following are several test items of the air and planning area performance test chamber to evaluate the purification effect of the air purifier.
1. Environmental condition control The air purifier test cabin has the simulation function to provide certain environmental conditions such as temperature, relative humidity and air exchange rate. When starting the air purifier test cabin, set the temperature, relative humidity and air flow required for the test. environmental conditions. After the system is running stably, within the next 48 hours, monitor the temperature, relative humidity and air exchange in the test simulation cabin near the exit.
2. The air purifier test cabin sealing test uses the tracer gas attenuation method. After the simulation cabin is sealed, the ambient temperature of the simulation cabin can be adjusted to 23°C and the relative humidity is 45%. In static mode, a certain amount of CO2 is injected into the air purifier test chamber, and based on the attenuation of the CO2 concentration in the chamber, a logarithmic-time relationship curve of CO2 concentration is drawn. The slope of the curve is the leakage amount.
3. The mixability of the simulation cabin can be quantitatively evaluated by the tracer gas attenuation method, using CO2 as the tracer gas and measuring it with a CO2 analyzer.
4. After the empty cabin operates stably and the concentration of some internal gases that may be related to the target object is low enough, it can be guaranteed not to interfere with the normal experiment. During the measurement, the simulation cabin was first cleaned with a high-flow pump for one night, and the air exchange volume was adjusted to 30m?/h. Then the background concentration was monitored at the intermediate sampling point and the outlet sampling point, and pollutants such as formaldehyde and VOCs were quantitatively detected.
Relevant test equipment and reagents: