This series of products are widely used in the aviation, aerospace, electronics, IT industries, instrumentation, electrical products and other simulated samples to test the adaptability of simulated samples under rapid changes in ambient atmospheric temperature and to provide reliability testing for the safety of electronic components. Product screening, etc.
GB/T2423.1-2008 Environmental testing of electrical and electronic products Part 2 Test methods Test A Low temperature
GB/T2423.2-2008 Environmental testing of electrical and electronic products Part 2: Test methods Test B: High temperature1. It adopts an industrial computer-type controller and is equipped with a large-area LED display and touch screen. It has powerful functions, strong scalability, and easy operation.
2. Two tanks (high temperature tank and low temperature tank) adopt a mobile sample hanging basket structure.
3. Five-minute recovery time, institution transfer time within 10 seconds.
4. The air volume is evenly distributed to ensure uniform test exposure.
5. Use the latest environmentally friendly refrigerant.
The two-box hot and cold shock box uses two upper and lower boxes (upper box with high temperature and lower box with low temperature). There is a movable basket in the middle. Samples can be placed in the movable basket. The basket can be switched up and down through the transmission device to quickly switch between the high and low temperature boxes. Thus achieving temperature shock.
1. Temperature and humidity measurement: Pt100 platinum resistor measures air temperature;
2. Interface language: Chinese/English can be switched at will;4. Main technical indicators | |
Cabinet: | |
1.Inner chamber size | About W800*H600*D700 mm |
2. Overall dimensions | About W1350*H2250*D2050mm (subject to the actual object) |
3. Hanging basket size | Approximately W600*H500*D500 mm hanging basket volume: 150L |
temperature: | |
1. High room temperature range | +50ºC~+180ºC |
2. Temperature rise time | R.T (room temperature)~+180ºC ≤30min |
3. Low temperature room temperature range | 0ºC~-70ºC |
4. Temperature drop time | R.T (room temperature)~-70ºC ≤50min |
Basket | |
1. Temperature shock range | -55ºC~+150ºC |
2.Temperature fluctuation | ≤±0.5ºC (no load, constant state) |
3. Temperature uniformity | ≤2ºC (no load, constant state) |
4. Temperature deviation | ≤±2ºC (no load, constant state) |
5.Conversion time | ≤10S (no load) |
6.Test conditions | High temperature +120ºC, exposure time 30min; Low temperature -55 ºC, exposure time 30min |
7. Temperature recovery time | Test area temperature recovery time: ≤4min. Impact process (as shown in the figure) |
8. Temperature shock method | The controller uses pneumatic actuators to control the cylinder to control the basket to quickly switch up and down between the low temperature box or the high temperature box, forming a closed-circuit air circulation system with the high temperature box or low temperature box respectively to quickly reach the test target temperature. |
9. Noise | Within 75db (measured 1 meter from the front of the machine and 1.2 meters from the ground) |
Note: The above performance indicators are values measured under the conditions of ambient temperature +25ºC, relative humidity ≤85%, and no sample. |