Walk-in thermal shock chamber is a customized three zones temperature shock chamber with a static test chamber to ensure that the sample will not be affected by vibration. The walk-in thermal shock chamber can be used to test different materials and is an essential equipment for many laboratories. The structure can be customized according to the site and test requirements.
GB/T2423.1-2008 Environmental testing of electrical and electronic products Part 2 Test methods Test A Low temperature
GB/T2423.2-2008 Environmental testing of electrical and electronic products Part 2: Test methods Test B: High temperatureTemperature range: -70ºC~+200ºC
Water-cooled cooling method
Color touch screen
Insulated glass window
Cable port
Shelves
Movable casters
Safety Protection Device
Heater empty ignition prevention protection switch
Heater overcurrent circuit breaker
Circulation fan over-current overload protection
Compressor high pressure protection switch
Compressor overheat protection switch
Compressor overcurrent protection switch
Line isolating switch
Leakage switch
Water shortage protection
Main Specifications | ||
Cabinet: | ||
Studio size | Approximately W 500 × H 450 × D 400 (mm) (W width × H height × D depth) | |
Dimensions | Approximately 1450×1800×1650(mm), subject to the actual object | |
Volume | 100L; | |
Noise | Within 75dB (measured 1 meter from the front of the machine and 1.2 meters from the ground) | |
Temperature: | ||
High temperature zone |
Temperature range | RTºC~+200ºC |
Heating time | RT~+200ºC: about 30min | |
Low temperature zone |
Temperature range | RT~-55ºC |
Cooling time | RT~-55ºC: about55min | |
Studio |
Temperature shock range | -65ºC~+150ºC |
Temperature fluctuation | ≤±0.5 | |
Temperature deviation | ≤±2ºC | |
Temperature uniformity | ≤2ºC | |
Temperature conversion time | ≤10s(Damper opening time) | |
Temperature recovery time | ≤4min | |
Operation mode | There are two high and low temperature impact modes and high and low temperature + room temperature impact modes, which can be set by the human-machine interface. | |
High and low temperature shock mode | 150ºC(30min)-NC--40ºC(30min) | |
High and low temperature + room temperature impact mode | 150ºC(30min)-room temperature--40ºC(30min) | |
Power | High greenhouse power: 9KW Low temperature greenhouse power: 9KW Other power about: 2KW Maximum power about: 20KW Operating power approximately: 12KW |
|
Voltage | Power requirements: 380V 50HZ three-phase five-wire system plus ground wire Power: 20KW Current 40A |
|
Condensation method | Water cooling |
1.Q :How to provide after-sales service?
A:If customers need our engineers to come to the site for service, we will send out engineers to provide repair service for you.
2.Q: Can we visit and study in your factory?
A: Welcome to visit and study in our factory, we will show you our R&D and production capacity.
3.Q :Can I install and commission the calibration equipment on site?
A: Yes, we can send technicians and engineers to the customer's site for installation, commissioning and calibration.
4.Q: What are the transportation methods?
A :Air freight, sea freight, railroad transportation, all transportation methods according to the customer's needs.
5.Q : What are the packing methods?
A: Standard export wooden box packaging, solid and firm.
6.Q :How do you calibrate the equipment?
A: We have our own calibration and third-party calibration.
7.Q :What are the payment methods?
A :The payment methods are T/T, CIF, Paypal, L/C, VISA and so on.
8.Q :How long is the warranty period?
A :Warranty is 1 year, lifetime free service.
9.Q :What is the delivery time for your company's products?
A :This depends on the actual product you order, usually at least 30 days.