Customization: | Available |
---|---|
After-sales Service: | 1 Year |
Power Supply: | AC 380V |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Walk-in thermal shock chamber is a customized three zones temperature shock chamber with a static test chamber to ensure that the sample will not be affected by vibration. The walk-in thermal shock chamber can be used to test different materials and is an essential equipment for many laboratories. The structure can be customized according to the site and test requirements.
GB/T2423.1-2008 Environmental testing of electrical and electronic products Part 2 Test methods Test A Low temperature
GB/T2423.2-2008 Environmental testing of electrical and electronic products Part 2: Test methods Test B: High temperatureTemperature range: -70ºC~+200ºC
Water-cooled cooling method
Color touch screen
Insulated glass window
Cable port
Shelves
Movable casters
Safety Protection Device
Heater empty ignition prevention protection switch
Heater overcurrent circuit breaker
Circulation fan over-current overload protection
Compressor high pressure protection switch
Compressor overheat protection switch
Compressor overcurrent protection switch
Line isolating switch
Leakage switch
Water shortage protection
Main Specifications | ||
Cabinet: | ||
Studio size | Approximately W 500 × H 450 × D 400 (mm) (W width × H height × D depth) | |
Dimensions | Approximately 1450×1800×1650(mm), subject to the actual object | |
Volume | 100L; | |
Noise | Within 75dB (measured 1 meter from the front of the machine and 1.2 meters from the ground) | |
Temperature: | ||
High temperature zone |
Temperature range | RTºC~+200ºC |
Heating time | RT~+200ºC: about 30min | |
Low temperature zone |
Temperature range | RT~-55ºC |
Cooling time | RT~-55ºC: about55min | |
Studio |
Temperature shock range | -65ºC~+150ºC |
Temperature fluctuation | ≤±0.5 | |
Temperature deviation | ≤±2ºC | |
Temperature uniformity | ≤2ºC | |
Temperature conversion time | ≤10s(Damper opening time) | |
Temperature recovery time | ≤4min | |
Operation mode | There are two high and low temperature impact modes and high and low temperature + room temperature impact modes, which can be set by the human-machine interface. | |
High and low temperature shock mode | 150ºC(30min)-NC--40ºC(30min) | |
High and low temperature + room temperature impact mode | 150ºC(30min)-room temperature--40ºC(30min) | |
Power | High greenhouse power: 9KW Low temperature greenhouse power: 9KW Other power about: 2KW Maximum power about: 20KW Operating power approximately: 12KW |
|
Voltage | Power requirements: 380V 50HZ three-phase five-wire system plus ground wire Power: 20KW Current 40A |
|
Condensation method | Water cooling |