VOC formaldehyde release test chamber is suitable for volatile organic compound release sample equipment, used to determine volatile organic compound release,
total volatile organic compound (TVOC), benzene, toluene, xylene and ethylbenzene sum, carbon disulfide and formaldehyde and other harmful substances release test.
Formaldehyde, VOC (Volatile Organic Compounds), benzene, toluene and other components in TVOC are strong carcinogens recognized by the World Health Organization.
Long-term exposure to pollutants with excessive concentrations will seriously endanger people's health. The amount of pollutants released is tested and limited.
will help improve product quality and protect people's health.
Wood products, furniture, floors, carpets, footwear products, building decoration materials, automobile interiors and other materials that come into contact
with the human body will emit VOC (volatile organic compounds), formaldehyde and other harmful substances to the human body,
especially for If products are placed densely and in closed spaces indoors or in cars, the concentration will be higher after accumulation, which will be more harmful to human health.
This test chamber is an instrument and equipment used to simulate the release of VOC (formaldehyde) from products in an environment close to where people live.
It is used to test and evaluate the amount of harmful substances released in furniture, building materials, automobile interiors and other products.
Standard basis:
GB18580-2017 "Indoor Decoration Materials - Formaldehyde Release Limits in Artificial Panels and Their Products"
GB18584-2011 "Indoor Decoration Materials-Limits of Hazardous Substances in Wooden Furniture" (under review)
Appendix A (Normative Appendix) in GB18587-2001 "Indoor Decoration and Renovation Materials-Carpets, Carpet Pads
and Carpet Adhesives Release Limits of Hazardous Substances": Small Environmental Test Chamber Method.
GB50325-2010 "Code for Indoor Environmental Pollution Control of Civil Construction Projects"
GB/T18584 General technical conditions for climate chambers for detection of volatile organic compounds in furniture (draft for approval)
GB/T28489-2012 "Limits of Hazardous Substances in Musical Instruments"
GB/T29592-2013, GB/T29592-2013 "Method for determination of release of volatile organic compounds (VOC) and aldehydes from construction adhesives"
GB/T29899-2013<<Test method for release of volatile organic compounds in artificial panels and their products - Small release chamber method>>
LY/T1980-2011 "Volatile organic compounds and formaldehyde release detection box"
LY/T1981-2011 "Formaldehyde Release Gas Analysis Detection Box"
JG/T344-2011 "Indoor Environmental Test Chamber for Construction Engineering"
HJ571-2010 "Technical Requirements for Environmental Labeling Products Artificial Panels and Their Products"
HJ/T201-2005 "Technical Requirements for Environmental Labeling Products Water-based Coatings"
HJ/T220-2005 "Technical Requirements for Environmental Labeling Products Adhesives"
HJ/T414-2007 "Technical Requirements for Environmental Labeling Products Solvent-based Wood Coatings for Interior Decoration"
International standards:
ISO16000-9-2006 "Determination of volatile organic compounds released from building products and furniture - Laboratory release method"
Temperature
Temperature range +10 ~ 50 ºC (no load during test operation, when there is a sample)
Temperature fluctuation ±0.5 ºC (no load, no sample)
Temperature uniformity ≤±0.5 ºC (no load, no sample)
Temperature resolution 0.1 ºC
Temperature control: Air jacket indirect temperature control method (heating and cooling components are externally installed)
Humidity
Humidity range 30~90% RH (no load, with sample)
Humidity fluctuation ≤ ± 3% RH (no load, no sample)
Humidity uniformity ≤ ±2 % RH (no load, no sample)
Humidity resolution 0.1 % RH
Humidity control: dry and wet gas proportion control method (external), stabilization time ≤50 min
Background value
Formaldehyde background value <0.006 mg/m3
Single VOC background <0.002 mg/m3
TVOC background value <0.02 mg/m3
Air exchange rate (0.2~2) times/h, control accuracy ±3%, display 0.01 times/h
Air and seal programmable settings
Center wind speed 0.1~1.0 m/s (adjustable)
Air flow velocity (0.1~0.5) m/s, display accuracy ±0.05m/s
Sealing performance: Add 1000Pa overpressure, gas leakage is less than 1×10-3m3/min
Equipment recycling rate:
>85% or more, the recovery rate test contains at least three standard substances toluene
and n-dodecane, the polar compound tetraphenylcyclohexene
Relative positive pressure maintained at 1000±5 Pa
Display accuracy required for cabin environment:
1. Temperature 0.1ºC; 2. Relative humidity 1%; 3. Air flow rate;
Material
Cabin material
Imported SUS 304 mirror stainless steel 2 mm, corners rounded, fully welded and polished
Light technology
Shell mat
Technical parameter:
Studio volume |
1000L |
Studio |
Customized |
Exterior |
Customized |
Temperature range |
15~70ºC ★Meet the requirements of plastic surface testing(60ºC,5%R.H.) |
Temperature fluctuation |
≤ ±0.5ºC |
Temperature uniformity |
≤±0.5ºC |
Temperature resolution |
0.1ºC |
Temperature control |
Indirect temperature control method in constant temperature room |
Humidity range |
3~80 % R.H. ★Meet the requirements of plastic surface testing(60ºC,5%R.H.) |
Humidity Fluctuation |
≤ ±2% RH |
Humidity uniformity |
≤ ±2 % RH |
Humidity Resolution |
0.1 % RH |
Humidity control |
Dry and moisture proportional control method (external) |
Formaldehyde background value |
≤0.006 mg/m3 |
Individual VOC background value |
≤0.002 mg/m3 |
TVOC background value |
≤0.02 mg/m3 |